FDA recall Z-2031-2025

Baxter Healthcare Corporation · Class I · device

Product

Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift - Product Code 2040045A; 5) Viking S Mobile Lift - Product Code 2040006; 6) Viking XS Mobile Lift - Product Code 2040007; 7) LikoLight Mobile lift - Product Code 2030001. .

Reason for recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Distribution

Worldwide

Key facts

Status
Ongoing
Initiation date
2025-05-30
Report date
2025-07-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2031-2025