## Product

Hillrom LikoScale component, Product Code 3156232,  used with the following adapter kits for patient lifts:    1) LikoScale 200 Accessory - Product Code 3156225;   2) LikoScale 350 Accessory - Product Code 3156228;   3)  LikoScale 400 Accessory - Product Code 3156226.

## Reason for recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

## Distribution

Worldwide

## Key facts

- **Recall number:** Z-2032-2025
- **Recalling firm:** Baxter Healthcare Corporation
- **Classification:** Class I
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-05-30
- **Report date:** 2025-07-16

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Deerfield, IL, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2032-2025