FDA recall Z-2041-2023

Phadia US Inc · Class II · device

Product

Phadia 2500EE instrument -Intended for the in vitro semi-quantitative measurement of IgG antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease Material Number 12410002

Reason for recall

Higher reported results when running EliA GliadinDP IgG Well on the instruments Phadia 2500EE compared to running EliA GliadinDP IgG Well on Phadia 250 instrument

Distribution

US: CA, NJ, MI

Key facts

Status
Ongoing
Initiation date
2023-05-01
Report date
2023-07-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portage, MI, United States

Primary source