FDA recall Z-2041-2026

Medline Industries, LP · Class II · device

Product

Medline and Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. FOLEY CARE KIT, Medline Kit SKU UROT1044; 2. IV START KIT WITH EXTENSION TUBING, Medline Kit SKU IV8770; 3. IV START KIT, Medline Kit SKU DYNDV2520; 4. ARTERIAL LINE BUNDLE, Medline Kit SKU ART255; 5. ON/OFF TRAY, Medline Kit SKU DYNDC2969; 6. IV START KIT, Medline Kit SKU DYNDV2520; 7. IV START KIT WITH EXTENSION TUBING, Medline Kit SKU IV8770; 8. (Centurion) ARTERIAL LINE BUNDLE, Medline Kit SKU ART255; 9. ON/OFF TRAY, Medline Kit SKU DYNDC2969; 10. CVAD DRESSING CHANGE KIT, Medline Kit SKU EBSI1364A.

Reason for recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-03-16
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2041-2026