FDA recall Z-2045-2026

Medline Industries, LP · Class II · device

Product

Medline and Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. LVAD DRIVELINE KIT (Centurion), Medline Kit SKU DT17360 (Centurion); 2. LONG TERM/INPATIENT IV SECUREMENT SET (Centurion), Medline Kit SKU IVS1975A; 3. CVC DRESSING CHANGE TRAY (Centurion), Medline Kit SKU DT5565B; 4. BSI PICC DRESSING CHANGE KIT, Medline Kit SKU EBSI1172B; 5. BSI CVC DRESSING CHANGE KIT, Medline Kit SKU EBSI1171A; 6. BSI SENSITIVE SKIN DRSNG CHNGE, Medline Kit SKU EBSI1173A; 7. LVAD MANAGEMENT SYSTEM - DAILY (Centurion), Medline Kit SKU DM810; 8. WEEKLY LVAD TRAY (Centurion), Medline Kit SKU DM100; 9. DRESSING CHANGE TRAY (Centurion), Medline Kit SKU DT7340; 10. SENSITIVE SKIN LVAD TRAY (Centurion), Medline Kit SKU DM660; 11. PORT DRESSING CHANGE KIT, Medline Kit SKU EBSI1365A.

Reason for recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-03-16
Report date
2026-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2045-2026