# FDA recall Z-2053-2020

> **Philips North  America, LLC** · Class II · device recall initiated 2020-04-13.

## Product

Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.

## Reason for recall

The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.

## Distribution

Worldwide distribution - US Nationwide distribution.

## Key facts

- **Recall number:** Z-2053-2020
- **Recalling firm:** Philips North  America, LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2020-04-13
- **Report date:** 2020-05-27
- **Termination date:** 2021-04-07

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Andover, MA, United States