FDA recall Z-2054-2020

Skeletal Kinetics, Llc · Class II · device

Product

Cranio/Sculpt C, 10cc, Bone Void Filler, UDI: 00813845021129 - Product Usage: CRANIO/SCULPT C is a calcium phosphate bone void filler indicated for the repair or filling of neurosurgical burr holes, other cranial bone defects and craniotomy cuts with a surface area no larger than 25cm2. CRANIO/SCULPT C may be used in the restoration or augmentation of bony contours of the cranial bone skeleton.

Reason for recall

Potential product mix-up. The recalling firm has determined that one unit of OsteoVation EX, 3cc that may not have been sterilized was inadvertently packaged as CranioSculpt C, 10cc.

Distribution

US Nationwide distribution in the state of Florida.

Key facts

Status
Terminated
Initiation date
2020-04-03
Report date
2020-05-27
Termination date
2025-01-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Addison, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2054-2020