FDA recall Z-2058-2019

Zyno Medical LLC · Class II · device

Product

Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072

Reason for recall

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Distribution

MA

Key facts

Status
Terminated
Initiation date
2019-05-29
Report date
2019-07-31
Termination date
2020-06-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Natick, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2058-2019