FDA recall Z-2062-2020

Innovasis, Inc · Class II · device

Product

Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usage: The Cervical Plate System is intended for use in anterior cervical fixation.

Reason for recall

Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw.

Distribution

US Nationwide distribution including in the states of CA, AL, LA.

Key facts

Status
Terminated
Initiation date
2020-04-07
Report date
2020-05-27
Termination date
2022-01-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Salt Lake City, UT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2062-2020