FDA recall Z-2065-2026

Physio-Control, Inc. · Class II · device

Product

LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122

Reason for recall

Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

Distribution

US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.

Key facts

Status
Ongoing
Initiation date
2026-04-13
Report date
2026-05-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Redmond, WA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2065-2026