FDA recall Z-2066-2019

Medline Industries Inc · Class I · device

Product

Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the subject of this recall, and the recalled BD Alaris Pump Infusion Set (RES#82951) Product Usage: The Vyaire enFlow Disposable Cartridges contain a heating element that is used to help raise or maintain a patient's body temperature.

Reason for recall

The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.

Distribution

One distributor located in Illinois.

Key facts

Status
Terminated
Initiation date
2019-05-03
Report date
2019-08-21
Termination date
2021-03-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source