FDA recall Z-2080-2019

Bard Peripheral Vascular Inc · Class II · device

Product

POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE isp, 8F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 6F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; POWERPORT CLEARVUE Slim, 8F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit; NAUTILUS DELTA Tip Confirmation System (Includes: Netbook with Pre-loaded Software, Patient Module, and ECG Cable) Product Usage: Is an implantable access device designed to provide repeated access to the vascular system.

Reason for recall

Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with the NAUTILUS DELTA" TCS.

Distribution

US Nationwide Distribution in the states of FL, CO, WI, OH, IL, KS, AZ, WV, CT, PA, AL, MI and VA

Key facts

Status
Completed
Initiation date
2018-06-25
Report date
2019-07-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tempe, AZ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2080-2019