FDA recall Z-2081-2019

Xcision Medical Systems, LLC · Class II · device

Product

GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

Reason for recall

One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

Distribution

The products were distributed to the following US states: MD and TX.

Key facts

Status
Terminated
Initiation date
2019-07-03
Report date
2019-07-31
Termination date
2021-07-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Columbia, MD, United States