FDA recall Z-2082-2019

Medical Intelligence Medizintechnik Gmbh · Class II · device

Product

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

Reason for recall

Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.

Distribution

Worldwide Distribution - US Nationwide

Key facts

Status
Terminated
Initiation date
2019-07-18
Report date
2019-07-31
Termination date
2023-08-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Schwabmunchen, N/A, Germany