FDA recall Z-2087-2018

Tosoh Bioscience Inc · Class II · device

Product

ST-AIA PACK Prolactin; Part Number: 025255 Assay, Reproductive Hormone

Reason for recall

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

Distribution

Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.

Key facts

Status
Terminated
Initiation date
2018-03-05
Report date
2018-06-13
Termination date
2020-08-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Grove City, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2087-2018