See RES for complete list. Medline Convenience Kits: 1) OPEN HEART CDS, Model Number:CDS840015AQ; 2) OPEN HEART CDS, Model Number:CDS840015AQ; 3) OPEN HEART CDS, Model Number:CDS840023T; 4) OFF PUMP CABG CDS, Model Number:CDS840087AI; 5) OFF PUMP CABG CDS, Model Number:CDS840087AK; 6) OFF PUMP CABG CDS, Model Number:CDS840087AL; 7) OFF PUMP CABG CDS, Model Number:CDS840087AM; 8) OPEN HEART, Model Number:CDS840150S; 9) OPEN HEART, Model Number:CDS840150T;
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Worldwide - US Nationwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2093-2026