# FDA recall Z-2105-2025

> **Boston Scientific Corporation** · Class II · device recall initiated 2025-06-10.

## Product

VersaCross Transseptal Sheath Kit (VXS)  Product Description/IDs  VXS0201  VXS 63CM/45DEG/D0 SHEATH    VXS2112  VERSACROSS SHEATH (63CM 45DEG D0)-J    Used for the percutaneous introduction of various types of  cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal  perforation / puncture.

## Reason for recall

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

## Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

## Key facts

- **Recall number:** Z-2105-2025
- **Recalling firm:** Boston Scientific Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-06-10
- **Report date:** 2025-07-16

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Saint Paul, MN, United States