FDA recall Z-2106-2020

Medtronic Inc. · Class II · device

Product

Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.

Reason for recall

Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2019-05-20
Report date
2020-06-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northridge, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2106-2020