FDA recall Z-2107-2020

Quest Medical, Inc. · Class II · device

Product

Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The device is a component of the MPS Console used as replacements for worn or damaged MPS Console top covers)

Reason for recall

Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention.

Distribution

US Nationwide distribution including in the states of NY, PA, TN, WV. OUS: None

Key facts

Status
Terminated
Initiation date
2019-12-17
Report date
2020-06-03
Termination date
2024-04-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Allen, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2107-2020