FDA recall Z-2110-2020

The Binding Site Group, Ltd. · Class II · device

Product

Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)

Reason for recall

Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.

Distribution

US: AL, AR, AZ, FL, GA, KY, MA, MD, MI, MN, MS, NC, NE, NY, OR, PA, TX, VT, WA, WI OUS: None

Key facts

Status
Terminated
Initiation date
2019-10-07
Report date
2020-06-03
Termination date
2022-01-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Birmingham, N/A, United Kingdom

Primary source