FDA recall Z-2114-2026

Medline Industries, LP · Class II · device

Product

See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W;

Reason for recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2026-01-07
Report date
2026-05-13
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2114-2026