FDA recall Z-2115-2021

Exactech, Inc. · Class II · device

Product

Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 130-22-XX, Novation GXL Neutral Liner, G00, 22mm ID. Orthopedic hip implant component.

Reason for recall

Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.

Distribution

Distribution US nationwide.

Key facts

Status
Ongoing
Initiation date
2021-06-29
Report date
2021-07-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Gainesville, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2115-2021