# FDA recall Z-2121-2025

> **Ortho-Clinical Diagnostics, Inc.** · Class II · device recall initiated 2025-06-10.

## Product

Brand Name: VITROS Chemistry Products  Product Name: VITROS Chemistry Products Performance Verifier II  Model/Catalog Number: 8231474  Software Version: N/A  Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems.    VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added.  Component: No, not a component

## Reason for recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

## Distribution

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

## Key facts

- **Recall number:** Z-2121-2025
- **Recalling firm:** Ortho-Clinical Diagnostics, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-06-10
- **Report date:** 2025-07-23

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Rochester, NY, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2121-2025

## Citation

> AI Analytics. FDA recall Z-2121-2025. Retrieved 2026-06-25 from https://api.ai-analytics.org/recall/Z-2121-2025. Source: US FDA. Licensed CC0.

---