FDA recall Z-2127-2020

Siemens Medical Solutions USA, Inc · Class II · device

Product

Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Reason for recall

The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.

Distribution

US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.

Key facts

Status
Terminated
Initiation date
2020-04-17
Report date
2020-06-03
Termination date
2020-08-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States