FDA recall Z-2128-2020

Siemens Medical Solutions USA, Inc · Class II · device

Product

Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

Reason for recall

The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.

Distribution

US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.

Key facts

Status
Terminated
Initiation date
2020-04-17
Report date
2020-06-03
Termination date
2020-08-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States