FDA recall Z-2130-2023

Stryker Medical Division of Stryker Corporation · Class II · device

Product

Power-PRO 2- A powered ambulance cot that consists of a platform, including a mattress, mounted on a wheeled, retractable X-frame that is designed to support and transport a maximum weight of 700 lb (318 kg). Model Number: 6507

Reason for recall

Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Switzerland, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2023-05-24
Report date
2023-07-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portage, MI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2130-2023