FDA recall Z-2132-2018

Orthofix, Inc · Class II · device

Product

Straight Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Holes (Part Number): a) 3 Hole (003-D-01001); b) 4 Hole (003-D-02001); c) 6 Hole (003-D-03001)

Reason for recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Key facts

Status
Terminated
Initiation date
2018-04-06
Report date
2018-06-13
Termination date
2021-06-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lewisville, TX, United States

Primary source