FDA recall Z-2132-2020

Smith & Nephew, Inc. · Class II · device

Product

JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5, CATALOG NUMBER: REF 74035825 SIZE 6, CATALOG NUMBER: REF 74035826 SIZE 7, CATALOG NUMBER: REF 74035827 SIZE 8, CATALOG NUMBER: REF 74035828 SIZE 9, CATALOG NUMBER: REF 74035829 SIZE 10, CATALOG NUMBER: REF 74035830

Reason for recall

The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-01-28
Report date
2020-06-03
Termination date
2023-12-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Memphis, TN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2132-2020