FDA recall Z-2135-2020

MOTEK MEDICAL B.V. · Class II · device

Product

Body Weight Support (BWS) Light; Model No. RB011; a component of C-Mill, Grail, M-Gait - Product Usage: is designed to facilitate functional gait training of subjects with gait and/or balance impairment by providing dynamic partial vertical unloading during walking on a treadmill. The device will be used as an assistive device for gait deficiency rehabilitation and research purposes.

Reason for recall

The firm discovered a potential manufacturing deviation of the yoke of the Body Weight Support (BWS) Light that in rare occasions may result in the yoke detaching from the cable guide.

Distribution

Worldwide distribution - US Nationwide distributions including in the states of AZ, CA, FL, KY, MA, MI, NJ, NY, NC, and OH. The countries of Argentina, Austria, Belgium, China, Curacao, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Japan, Mexico, Netherlands, Russia, South Korea, Spain, Switzerland, and United Arab Emirates.

Key facts

Status
Ongoing
Initiation date
2019-07-29
Report date
2020-06-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Amsterdam, N/A, Netherlands

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2135-2020