FDA recall Z-2140-2018

Alere San Diego Inc. DBA Immunalysis Corporation · Class III · device

Product

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.

Reason for recall

The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.

Distribution

US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV

Key facts

Status
Terminated
Initiation date
2017-11-28
Report date
2018-06-20
Termination date
2021-01-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Pomona, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2140-2018