FDA recall Z-2140-2021

Maquet Medical Systems USA · Class II · device

Product

Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.

Reason for recall

There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

Distribution

Distribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.

Key facts

Status
Terminated
Initiation date
2021-07-14
Report date
2021-08-04
Termination date
2022-08-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States