FDA recall Z-2142-2023

Hamilton Medical, Inc. · Class I · device

Product

HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates

Reason for recall

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

Distribution

U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New York North Carolina North Dakota Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Vermont Virginia Washington West Virginia Wisconsin Wyoming Added 08/30/2023 - Puerto Rico O.U.S.: Bahamas, Japan and Mexico

Key facts

Status
Ongoing
Initiation date
2023-06-15
Report date
2023-08-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Reno, NV, United States