FDA recall Z-2168-2025

Medtronic, Inc. · Class II · device

Product

Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;

Reason for recall

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-06-18
Report date
2025-07-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mounds View, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2168-2025