FDA recall Z-2187-2018

Roche Diabetes Care, Inc. · Class II · device

Product

Accu-Chek Inform II Test Strip, Whole blood glucose test system, Catalog no. 05942861001 Product Usage: The Accu-Chek Inform II test strips are for use with the Accu-Chek Inform II meter to quantitatively measure glucose (sugar) in venous whole blood, arterial whole blood, neonatal heel stick, or fresh capillary whole blood samples drawn from the fingertips as an aid in monitoring the effectiveness of glucose control. The system is not for use in diagnosis or screening of diabetes mellitus, nor for testing neonate cord blood samples.

Reason for recall

The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.

Distribution

Worldwide Distribution - US Nationwide distribution. Foreign distribution to Argentina Armenia Australia Austria Azerbaijan Bahamas Bahrain Barbados Belgium Brazil Bulgaria Canada Chile China Colombia Costa Rica Denmark Dominican Republic El Salvador Finland Georgia Germany Ghana Greece Guatemala Honduras Hong Kong (SAR China) Hungary India Indonesia Israel Italy Jamaica Japan Jordan Ireland Kenya Korea Kuwait Lebanon Lithuania Luxembourg Malaysia Malta Mexico Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Poland Russian Federation New Zealand Saudi Arabia Serbia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan (Republic of China), Thailand, United Arab Emirates, United Kingdom, Vietnam, Portugal"

Key facts

Status
Terminated
Initiation date
2018-05-07
Report date
2018-06-20
Termination date
2020-01-16
Voluntary/Mandated
Voluntary: Firm initiated