FDA recall Z-2198-2021

Straumann USA LLC · Class II · device

Product

Straumann SCS Screwdriver, driver, for ratchet, short, L 21mm, stainless steel Article Number: 046.401. dental surgical instrument.

Reason for recall

Tip of the screwdriver does not have the correct configuration / profile therefore the screwdriver does not properly fit the opposing screw head and cannot pick up the screws, may result in injury

Distribution

US Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2021-05-27
Report date
2021-08-11
Termination date
2023-03-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2198-2021