Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ARTERIAL LINE START KIT , Model Number ART920; b) ULTRASOUND GUIDED PIV KIT , Model Number BSIPIV1005; c) ULTRASOUND GUIDED IV START KIT W/SHIELD , Model Number BSIPIV1006; d) PIV BSI KIT , Model Number BSIPIV1010; e) ULTRASOUND GUIDED PIV BUNDLE W/CHG DRSG , Model Number BSIPIV1011; f) PICC INSERTION TRAY , Model Number CVI3805; g) PEDIATRIC PICC INSERTION TRAY , Model Number CVI4450B; h) CENTRAL LINE INSERTION TRAY , Model Number DT19810; i) NEONATAL PICC INSERTION TRAY , Model Number DYNDA2509A; j) USG PIV KIT , Model Number DYNDV2536; k) ULTRASOUND GUIDED IV START KIT , Model Number DYNDV2547; l) USG PIV KIT , Model Number DYNDV2567; m) POLY MIDLINE KIT 4F SL , Model Number DYNJ68937A; n) PICC KIT 4F SL PL MAX BARRIER , Model Number DYNJ70141MB; o) PICC KIT 5F DL PL MB , Model Number DYNJ70152MB; p) MIDLINE KIT 4F SL PL MAX BARR , Model Number DYNJ80141MB; q) MI
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2200-2023