FDA recall Z-2213-2024

Mermaid Medical A/S · Class II · device

Product

The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.

Reason for recall

Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.

Distribution

Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.

Key facts

Status
Ongoing
Initiation date
2024-05-01
Report date
2024-07-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Stenlose, Denmark

Primary source