FDA recall Z-2219-2019

Ideal Implant Incorporated · Class II · device

Product

IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.

Reason for recall

The implants were deflating due to a piece of silicone disrupting the seal of the posterior valve shortly after implantation.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Country of Canada.

Key facts

Status
Terminated
Initiation date
2019-03-18
Report date
2019-08-14
Termination date
2021-03-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Dallas, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2219-2019