FDA recall Z-2227-2018

The Seaberg Company Inc · Class II · device

Product

Part #: SAM XT-C, Model: Hi-Viz Orange or Civilian

Reason for recall

Based on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial blood flow. There is a remote chance that a serious injury and/or death could occur as a result of the failure of the device.

Distribution

USA: CA, MD, OH, FL, ID, CO, CT, OR, AZ, IL, ME, MO, NY, SC, MI, AR, MT, CT, WA, TX, KY, NC, NJ, PA OUS: Australia, Canada, China, Columbia, Costa Rica, Europe (EU), Hong Kong, Iceland, India, Iraq, Israel, Japan, Kenya, Korea, Lebanon, Malaysia, Russia, Saudi Arabia, Singapore, South Africa, Thailand, UAE

Key facts

Status
Terminated
Initiation date
2018-05-02
Report date
2018-06-20
Termination date
2020-07-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wilsonville, OR, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2227-2018