FDA recall Z-2227-2019

Edwards Lifesciences, LLC · Class I · device

Product

Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System (SAPIEN 3 Ultra Delivery System) (29mm) 9630TF29 UDI:00690103201260

Reason for recall

The action has been initiated to address reports of burst balloons which have resulted in significant difficulty retrieving the device into the sheath and withdrawing the system from the patient during procedures

Distribution

US: VA, CA, GA, MA, MD, NY, OR, WA, MN, TN, AL, KS OUS: United Kingdom, Switzerland, Sweden, Spain, Portugal, Poland, Norway, Netherlands, Italy, Ireland, Germany, Finland, Denmark, Canary Islands, Canada, Austria

Key facts

Status
Terminated
Initiation date
2019-07-09
Report date
2019-08-28
Termination date
2020-09-15