FDA recall Z-2236-2024

Exactech, Inc. · Class II · device

Product

Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R; f) 100-22-24, SZ. S; g) 100-22-25, SZ. T

Reason for recall

The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.

Distribution

Worldwide - US distribution in Puerto Rico and the countries of Great Britain, Japan.

Key facts

Status
Ongoing
Initiation date
2024-04-26
Report date
2024-07-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Gainesville, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2236-2024