FDA recall Z-2244-2021

Cardinal Health 200, LLC · Class I · device

Product

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Reason for recall

The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2021-06-16
Report date
2021-08-25
Termination date
2023-01-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mansfield, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2244-2021