# FDA recall Z-2247-2023

> **Johnson & Johnson Surgical Vision, Inc.** · Class II · device recall initiated 2023-05-30.

## Product

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

## Reason for recall

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

## Distribution

Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Korea¿(South), Kuwait, Lebanon, Malaysia, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.

## Key facts

- **Recall number:** Z-2247-2023
- **Recalling firm:** Johnson & Johnson Surgical Vision, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-05-30
- **Report date:** 2023-08-09

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Irvine, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2247-2023

## Citation

> AI Analytics. FDA recall Z-2247-2023. Retrieved 2026-07-06 from https://api.ai-analytics.org/recall/Z-2247-2023. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*
