FDA recall Z-2251-2020

Abbott Laboratories · Class II · device

Product

Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.

Reason for recall

In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of NC, ND, SC, TN. The countries of United Kingdom, France, New Zealand, Pakistan, Turkey.

Key facts

Status
Terminated
Initiation date
2020-04-02
Report date
2020-06-10
Termination date
2021-04-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irving, TX, United States