BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYXIS MB TWR MN 2FH-1FM-4HH-2FM MP/ 169-145, BD PYXIS MB TWR MN 2FM-8HH-1FH MP/ 169-144, BD PYXIS MB TWR MN 4FH-4HH-1FM MP/ 169-142, BD PYXIS MB TWR RXNOW MAIN W/MED PRINTER / 169-155, BD PYXIS MB TWR RXNOW MAIN W/PAT PRINTER/ 169-156, BD PYXIS MDBANK TWR MN 7HH-1HM-3FM/ 169-132, BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow/ 169-131, BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM/ 169-23, BD PYXIS MEDBANK TWR MN 10HH-2FH/ 169-24, BD PYXIS MEDBANK TWR MN 10HH-2FM/ 169-25, BD PYXIS MEDBANK TWR MN 12HH-1FF/ 169-26, BD PYXIS MEDBANK TWR MN 12HH-1FH/ 169-27, BD PYXIS MEDBANK TWR MN 12HH-1FM/ 169-28, BD PYXIS MEDBANK TWR MN 14HH/ 169-29, BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM/ 169-138, BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF/ 169-30, BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM/ 169-31, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/ 169-32, BD PYXIS ME
Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.
US Nationwide distribution including in the states of NY, CA, MO, AZ, CO, KS, IL, FL, OK, OH, TN, GA, NM, NV, NE, TX, NC, KY, WI, SC, IA, MT, SD, VA, NJ, PA, AK, MS, MI, WA, UT, ND, MA, MN, OR, IN, WV, AL, ID, LA, CT, WY, MD, NH, AR, HI, DC.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2256-2024