# FDA recall Z-2258-2024

> **GE Medical Systems Information Technologies Inc** · Class II · device recall initiated 2024-04-22.

## Product

Prucka 3 Amplifier Model #P1801PA used with CardioLab AltiX electrophysiology recording system and ComboLab AltiX hemodynamic and electrophysiology recording system; and Field Replacement Units spare part #5875569 (Assy CLABIII AMP 128CH 100-240V 50-60HZ).

## Reason for recall

Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.

## Distribution

Worldwide - US Nationwide distribution in the states of AK, AZ, CA, CO, FL, GA, IL, KY, MA, ME, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. The countries of Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Japan, Republic of Korea, and Singapore.

## Key facts

- **Recall number:** Z-2258-2024
- **Recalling firm:** GE Medical Systems Information Technologies Inc
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-04-22
- **Report date:** 2024-07-10

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Milwaukee, WI, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2258-2024

## Citation

> AI Analytics. FDA recall Z-2258-2024. Retrieved 2026-08-11 from https://api.ai-analytics.org/recall/Z-2258-2024. Source: US FDA. Licensed CC0.

---