Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Malaysia, Mexico, New Zealand, Taiwan, Thailand, Canada, Austria, Belgium, Cyprus, Czech Republic, Germany, Hungary, Ireland, Italy, Switzerland, Luxembourg, Netherlands, Poland, Portugal, Romania, Spain, Sweden, and United Kingdom