AHAR50K STANDARD ARTHROSCOPY PACK - 291852. Procedure tray Catalog Number: AHAR50K
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
US Nationwide distribution.