MAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23B
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
US Nationwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2276-2024