FDA recall Z-2279-2019

Becton Dickinson & Company · Class II · device

Product

BD Vacutainer Fluoride Tubes for Blood Alcohol determinations Part Number: 367001 Product Usage: Blood Alcohol determinations

Reason for recall

Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no longer than two days. If the sample was stored for more than 2 days, the result for blood alcohol determination might not be accurate (either falsely low or falsely high)

Distribution

US Nationwide-Distribution in the states of CA, FL, GA, IL, KS, MO, PA, RI, TN Foreign: Canada

Key facts

Status
Terminated
Initiation date
2019-05-30
Report date
2019-08-21
Termination date
2020-05-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Franklin Lakes, NJ, United States

Primary source